Comparative study of two automated pre-transfusion testing systems (microplate and gel column methods) with standard tube technique
Takako Ono1, Hitoshi Ohto1, Hiroyasu Yasuda1, Rie Hikichi1, Kinuyo Kawabata1, Keiji Minakawa1, Satoshi Ono1, Masami Kikuchi1, Akiko Sugawara1, Shunnichi Saito1, Nozomi Takano1, Kenneth E. Nollet1
1Department of Blood Transfusion and Transplantation Immunology, Fukushima Medical University Hospital, Fukushima, Japan

Article ID: 100030IJBTITO2017
doi:10.5348/ijbti-2017-30-OA-3

Address correspondence to:
Hitoshi Ohto
MD, PhD, Department of Blood Transfusion and Transplantation Immunology
Fukushima Medical University Hospital, Hikariga-oka, Fukushima City
Fukushima 960-1295 Japan

Access full text article on other devices

  Access PDF of article on other devices

[HTML Abstract]   [PDF Full Text] [Print This Article]
[Similar article in Pumed] [Similar article in Google Scholar]

How to cite this article:
Ono T, Ohto H, Yasuda H, Hikichi R, Kawabata K, Minakawa K, Ono S, Kikuchi M, Sugawara A, Saito S, Takano N, Nollet KE. Comparative study of two automated pre-transfusion testing systems (microplate and gel column methods) with standard tube technique. Int J Blood Transfus Immunohematol 2017;7:15–25.


ABSTRACT
Aims: Clinically significant antibodies may become undetectable and still provoke transfusion reactions. Hemolysis has been reported among transfusion recipients with anti-Kidd that was undetectable by gel column but detectable by other methods.
Methods: We compared two automated technologies - microplate (MP) and gel column (GC) methods - with manual methods in ABO/RhD typing, irregular antibody screening, identification, titration, and detection threshold of mixed-field agglutination. Results: Automated systems agreed generally with tube results in 98% or more of ABO forward and RhD groupings, but showed weaker reactions in ABO reverse testing against A1 (K=0.88 with MP and K=0.77 with GC) and B cells (K=0.66 with MP and K=0.68 with GC) and failed to detect some anti-A (2 of 273 samples with MP) and anti-B (2 of 273 with MP and 1 of 272 with GC). MP missed 2 (anti-E and –Fyb) of 8 antibodies and GC missed 5 (2 anti-E and 1 each of -Fyb, -Jka and –Lea) of 10 antibodies, which manual PEG-IAT detected. Among 11 known alloantibodies, MP detected 7 antibodies at higher dilution than tube PEG-IAT, whereas GC showed lower scores in 7 samples than tube PEG-IAT and missed 2 clinically significant antibodies (anti-C and anti-Fyb).
Conclusion: Automated systems, comparable to manual tube technique for forward grouping of ABO and RhD, are less sensitive in ABO reverse testing. As GC more often failed to detect clinically significant antibodies of low-titer, users should be especially mindful of possible post-transfusion hemolysis.

Keywords: ABO grouping, Gel column method, Irregular antibody, Microplate method, Tube technique


INTRODUCTION

Automated pre-transfusion testing has gained favor to overcome drawbacks of tube technique, which is subject to human error even in laboratories well-staffed with dedicated transfusion specialists. Studies comparing standard tube technique with automated pre-transfusion testing have focused on the sensitivity of irregular antibody detection [1][2][3][4][5]. Automated microplate (MP) systems have been reported to be more sensitive than standard tube technique [4]. Conversely, a collaborative cohort found gel column (GC) systems to be less sensitive than low-ionic strength solution-enhanced indirect antiglobulin tests in detecting clinically significant irregular antibodies [6] . In a recent series of 17 patients with post-transfusion hemolysis, anti-Jka was detectable by MP but not by GC [7]. In this study, we compare two desk-top automated systems, MP (ECHO, Immucor) and GC (WADiana Compact, Grifols), with standard tube technique for ABO grouping, RhD typing, irregular antibody screening, titration of known irregular antibodies, and detection threshold of mixed-field agglutination.


MATERIALS AND METHODS

Samples

To compare ABO/D blood grouping between a direct agglutination method on a microplate and a standard tube method, and to antibody screening between a solid phase (SP) method and a standard tube method, we processed 273 consecutive routine samples during October 2012. Then, to compare blood grouping and antibody screening between a gel column (GC) method and standard tube method, we processed 272 consecutive routine samples during December 2012. Samples of newborns were excluded from this study.


Standard tube method

For determining the ABO group of red blood cells (RBC), 1 drop of each reagent (Bioclone anti-A, Bioclone anti-B, Ortho-Clinical Diagnostic, Tokyo, Japan), and 1 drop of 3–5% patient’s RBCs were added to appropriately-labeled tubes, mixed, centrifuged, resuspended and examined for agglutination. For the ABO reverse test, 2 drops each of patient’s plasma and 1 drop of reagent cells (AFFIRMAGEN A1 or B cells, Ortho-Clinical Diagnostics) were mixed, centrifuged, resuspended and examined for agglutination.

For D typing, 1 drop of reagent (Bioclone anti-D, Ortho-Clinical Diagnostics) and 1 drop of the corresponding control (Rh-hr Control, Ortho-Clinical Diagnostics) and 3–5% patient RBCs were mixed, centrifuged, resuspended and compared for agglutination. If there was no agglutination of the RBCs in both the reagent and control tubes, the weak D test was performed using an indirect antiglobulin (IAT) procedure. The RBCs were incubated at 37oC for 30 minutes, then washed 3 times with saline. A rabbit-source antihuman reagent (Ortho Anti-IgG (Rabbit), Ortho-Clinical Diagnostics), was added to each tube, centrifuged, resuspended and examined for agglutination. IAT Control cells coated with IgG (prepared in-house) were added to each negative test to confirm antihuman reactivity.

The antibody screen test was performed using a, polyethylene-glycol (PEG)-IAT method. Two drops of patient’s plasma and 1 drop each of a 3-cell screen set (SURGISCREEN, Ortho-Clinical Diagnostics) and a special screen cell (Di(a+), Ortho-Clinical Diagnostics) were mixed, centrifuged, resuspended and examined for direct agglutination. To each tube, 2 drops of 20% PEG (prepared in-house)[8] were added as a enhancer and incubated at 37oC for 15 minutes, washed 4 times with saline, and 2 drops of antihuman reagent (Ortho Anti-IgG (Rabbit), Ortho-Clinical Diagnostic) were added. After incubation, and centrifugation, the tubes were examined for agglutination. The in-house IgG-coated RBCs were added to each negative test. [7]

If any patient sample reacted with any screen cell, antibody identification was performed using an eleven-cell panel (RESOLVE Panel A and B, Ortho-Clinical Diagnostics) and/or sixteen-cell panel (PANOCELL-16, Immucor K.K., Tokyo, Japan) with PEG-IAT.


Solid phase method using an automated instrument

For ABO/D grouping, performed by direct agglutination method on a microplate 20 µL of patient’s red cell suspension and 40 µL of murine monoclonal ABO reagents (anti-A, anti-B, Immucor K.K.,) and monoclonal Rh reagent and control (Gamma-clone anti-D and Monoclonal Control, Immucor K.K.) were pipetted onto a microwell plate (Galileo Echo, Immucor K.K.), centrifuged, resuspended and agglutination was interpreted by the automated algorithm and camera. For ABO reverse typing, 15 µL of each of the reverse grouping cells (Referencell, Immucor K.K.) and 30 µL of patient’s plasma were mixed, centrifuged, resuspended and examined for agglutination.

For the solid phase antibody screening, 25 µL of patient’s plasma and 50 µL of additive (Capture LISS, Immucor K.K.) were added to the plate wells containing the RBC monolayers (Capture-R Ready-Screen(3), Immucor K.K.), The automated analyser (Galileo Echo, Immucor K.K.) completed the testing, added the indicator cells (Capture-R Ready Indicator Cells) and interpreted the results. All negative reactions, were checked visually for confirmation per the Manufacturer’s Operators’ Manual.

When any potential antibody was suspected, identification was performed using the manual PEG-IAT method.


Gel column method using an automated instrument

For ABO/D grouping, 10 µL of patient’s RBC suspension was added to the appropriate column on the gel card. The gel cards are prepared with antisera (DG Gel ABO/Rh (2D) Card, Kainos Laboratories, Tokyo, Japan) already added. The ABO reverse test was performed using 50 µL of 0.8% RBC (DG Reverse-Cyte A1,B 0.8%, Kainos Laboratories) All pipetting and processing is performed by an automated analyzer (WADiana Compact, Kainos Laboratories). After centrifugation, the cells pellet either at the top (positive reaction) or bottom (negative reaction) of the matrix. The results are read automatically by the instrument and can be confirmed visually

For the antibody screen, 50 µL of patient’s plasma were mixed with 25 µL each of 3 screen cells (Search-Cyte TCS 0.8%, Kainos Laboratories), incubated at 37oC for 15 minutes, centrifuged and interpreted by the instrument. Reactions could be read by the instrument and interpreted visually.

Antibody identification, if needed, was performed using the manual PEG-IAT method.


Dilution detection sensitivity of antibodies

Antibody detection sensitivity of the three methods was compared using 11 frozen samples including one each of anti-D, -E, -C, -Jka, -Fyb, -Lea, -S, -Dia,-Jra, and two anti-M. In order to minimize the influence of freezing and thawing, frozen samples were thawed once at 37°C, divided into two aliquots, and cryopreserved again at -30°C until testing. After re-thawing, serially diluted samples were measured by two automated methods and by tube PEG-IAT. Subsequently, agglutination strengths and antibody titration was scored according to the AABB Technical Manual [9].

Part of this study, comparing MP with tube PEG-IAT, was a previous brief report [10].


Detection thresholds of mixed-field agglutination

Thresholds of mixed-field agglutination were compared among three methods by using the samples prepared with group A and group O red blood cells from healthy donors, adjusting the mixing ratios in 10% increments.


Statistical analysis

For ABO, RhD grouping and indirect antiglobulin screening for irregular antibodies, Kappa statistics quantified the agreement of reaction strength between the two automated systems and the tube technique. Kappa (K) values of K(0.81–1.0) indicate near perfect agreement; K(0.61–0.8), substantial agreement; K(0.41–0.6), moderate agreement; K(0.21–0.4), fair agreement; and K(0.0–0.2), slight agreement. Titration score was analyzed by the Friedman test and the nonparametric Wilcoxson’s signed rank test. Differences with a p value of 0.05 or less were considered statistically significant. All statistical analyses were performed with SPSS Statistics version 24 (IBM Corporation).


RESULTS

ABO grouping and RhD typing

In ABO forward grouping of 273 samples, MP concordance with standard tube technique was 99.3% (K = 0.985) for A and 98.9% (K = 0.976) for B in agglutination strength (Table 1A and Table 1B ). Four samples containing fibrin clots were indeterminate by MP. GC concordance was 100% (K = 1.0) both for A and B of 272 samples.

In ABO, reverse testing (Table 1A and Table 1B ), MP was concordant with tube technique in 273 samples at 93.8% (K = 0.884) for agglutination strength against A1 cells and 79.5% (K = 0.663) against B cells. Gel column method was concordant with tube technique in 272 samples at 86.1% (K = 0.769) against A1 cells and at 79.2% (K = 0.675) against B cells. MP missed 2 samples of anti-A and 2 samples of anti-B, and showed weaker reactions than tube technique in 54 samples of anti-B. Similarly, GC missed one anti-B, and had weaker reactions than tube technique in 30 anti-A and 54 anti-B. MP required re-examination and/or detailed examination due to indeterminate ABO reverse grouping in 14 (5.1%) samples; in contrast only four samples (1.5%) by tube technique needed such examinations. Four samples in GC were indeterminate due to slightly unclear blood cell borders, and one sample was false positive due to the influence of strong chyle. Gel column method required visual confirmation and re-examination for 7 (2.6%) samples; in contrast, tube technique had no indeterminate results.

In RhD typing (Table 1A) and (Table 1B), both MP and GC showed good agreement with tube technique: 98.9% (K = 0.75) and 99.6% (K = 0.80) concordance, respectively.


Irregular antibody screening

In routine irregular antibody screening, MP and GC methods showed substantial concordance with tube technique, at 98.9% (K = 0.79) and 97.8% (K = 0.61), respectively (Table 2). However, MP missed one anti-E and one anti-Fyb out of 8 alloantibodies, which tube PEG-IAT detected. Similarly, GC failed to discover 5 irregular antibodies (two anti-E and one each of anti-Fyb, -Jka, and -Lea) out of 10 clinically significant antibodies, which showed weak agglutination by tube PEG-IAT. One anti-E was ultimately detected by GC re-examination using enzyme (papain)-treated RBCs. Conversely, tube PEG IAT missed one autoantibody, which MP detected, and one cold-reactive anti-M, which GC detected.


Sensitivity of two automated systems and the tube technique in the detection of red cell alloantibodies

Microplate detected at higher dilutions than tube PEG-IAT and GC a total of six antibodies (anti-E, -Jka, -Jra, -Dia, -D, and anti-C) among 8 clinically significant irregular antibodies (MP versus PEG-IAT: p<0.05, MP versus GC: p<0.05); PEG-IAT was superior to GC (p<0.05) in identifying eight clinically significant antibodies. Anti-C, which MP and tube PEG-IAT detected, was missed by GC. However, anti-Fyb, which tube PEG-IAT detected, was missed by MP and GC. Among three clinically less significant antibodies, GC was most sensitive, detecting two anti-M samples; MP did not detect an anti-Lea that the other two methods detected (Figure 1).


Detection threshold of mixed-field agglutination

Tube technique was able to identify mixed-field agglutination with the best resolving power using artificially composed chimeras of group O and A1 cells in ratios of 1:9 through 9:1, whereas MP identified chimeras at 60–80%, but not when O cells comprised 10%, 50%, and 90% of a sample. Gel column method did not identify chimeras at 10–30% of group O cell composition, but identified them at 40–90% (Table 3).


Cursor on image to zoom/Click text to open image
Table 1A: Comparison of reaction strength between microplate method and tube technique in ABO forward/reverse grouping and RhD typing


Cursor on image to zoom/Click text to open image
Table 1B: Comparison of reaction strength between gel column and tube technique in ABO forward/reverse grouping and RhD typing


Cursor on image to zoom/Click text to open image
Table 2 Comparison of indirect antiglobulin test for irregular antibody screening


Cursor on image to zoom/Click text to open image
Figure 1: Six antibodies, anti-E, -Jka, -Jra, -Dia, -D, and -C, could be detected up to the highest dilution magnification by microplate method. The most sensitive antibodies by the PEG-IAT included anti-S, -Fyb, and -Lea. Anti-Fyb was detected only by the PEG-IAT. The most sensitive antibodies by gel column method were two anti-M samples. Anti-C was not detected by gel column method.

Cursor on image to zoom/Click text to open image
Table 3: Comparison in the detection of chimera/mixed-field agglutination with three methods



DISCUSSION

The tube technique and two automated systems showed good concordance, with rates of 98% in ABO forward grouping and RhD typing. Contamination with small fibrin clots caused indeterminate results in MP. In ABO reverse grouping, MP showed better agreement with tube technique than GC, but reaction strength in MP was weaker than that in tubes. Such slightly weaker reactions of GC at detecting anti-B iso-hemagglutinins have been reported [11]. Weaker reactions may originate from the technological principles differentiating the three methods. Gel column method subjects red blood cells to ordinary gravity over a period of time, so when antigen-antibody affinity is low, an antibody could dissociate and ultimately settle out on a bottom layer, even as red blood cells crosslink. On the other hand, tube technique forms pellets by centrifugal force, and can detect weak reactivity enhanced by breaking contact with the tube wall. Microplate dispenses samples and reagent, centrifuges after the reaction occurs, shakes (refloating and shaping the blood cells), and finally uses a reader to measure agglutination. The shaking operation is stronger than that of the tube technique, and may cause a weaker agglutination than that in tubes.

In addition, weaker reactivity to B blood cells than A1 blood cells could be explained by having fewer antigen-binding sites. The number of antigens expressed on A1 adult blood cells is about 1×106/cell versus about 0.7×106 for group B red blood cells [12]. This difference may explain why mis-transfusion of A1 red cells, compared with B cells, to group O recipients is more likely to be fatal [13]. It is also reported that the major causes of mismatch between the result of forward and reverse ABO blood grouping is caused by weak reactivity of anti-B [11]. Discrepancies arise from other causes, such as weak subgroups of A or B and ABO non-identical hematopoietic stem cell transplantation. Thus the numerical difference in expression of A and B antigen, and the difference in antibody titers between anti-A and anti-B in patient plasma are consistent with the current results.

In irregular antibody detection study for routine screening, MP missed 2 and GC missed 5 clinically significant antibodies, all of which tube PEG-IAT detected. Our modified PEG-IAT (2 drops PEG vs. 4 drops in original [9]) in PEG-IAT may have enhanced detection of weak antibodies, by avoiding over-dilution. Conversely, MP detected one autoantibody and GC detected one cold-reactive anti-M that tube PEG-IAT missed. It has been reported that warm-autoantibodies are more likely to be detected by solid phase versus column [2]. Column methods and solid phase methods usually use anti-IgG antibodies for their antiglobulin test, and they show low detection sensitivity for antibodies of low clinical significance, such as IgM-type antibodies [14][15], but there are cases of IgM-type antibodies being detected. By warming and washing in tube methods, IgM antibodies may dissociate from red blood cells, and form agglutination in a few cases [3]. One patient having anti-E required a blood transfusion before completion of antibody testing, but no acute hemolysis, delayed hemolytic transfusion reaction (DHTR), or delayed serological transfusion reaction (DSTR) was observed, possibly the transfused red cells may have been E-negative.

There have been reports of antibodies undetectable by automated systems [7] [16][17]. Anti-Jka, most notorious for transient detection and subsequent acute hemolytic transfusion reaction and clinically overt DHTR after transfusion of Jka-positive red blood cells, was significantly missed by GC in 30.5% (32/105) of patients, which a solid-phase adherence test could discover, and, importantly, 17 recipients developed DHTR (6), possible DHTR (3), or DSTR (8) following Jka antigen-unknown red cell transfusion [7] . Conversely, GC is reported to be comparable to tube PEG-IAT in irregular antibody detection and resulted in a lower incidence of DHTR and DSTR [5]. A fully automated solid-phase red cell adherence system (ABS 2000) is reported to be discordant with tube technique (using an albumin or low-ionic-strength saline additive) in 245 (3.3%) of 7510 samples; 19 manual method false-negatives (1 anti-Cw, 6 anti-E, 3 anti-a, and 7 warm autoantibodies), 9 solid-phase false-negatives (1 anti-D, 2 anti-K, 4 anti-M, and 2 anti-Lea), 2 manual method false-positives and 7 solid-phase false-positives [2]. Similarly, in the detection of irregular antibodies, two solid-phase microplate assays (Capture-R and Solidscreen) are reported to be more sensitive by 18–22% compared with PEG-IAT tube method [4].

The detection rates of patients with irregular red cell antibodies at our institute ranged between 113/10,000 by tube albumin-IAT and 125/10,000 by PEG-IAT [8]. In this survey, the detection rates were better than our historical records: 259/10,000 by MP, 220/10,000 by GC, and 333/10,000 by PEG-IAT. However, of note, antibodies of low strength but clinically significant were missed in 25% of cases by MP and 50% by GC.

Microplate generally showed higher AABB agglutination score by titration studies using frozen antibody-positive samples (Figure 1). Anti-Fyb with weak reactivity by tube PEG-IAT was not detectable by MP and GC. Anti-C failed to be detectable by GC. One of two anti-M samples was not detected by MP using Capture-R Ready-Screen (III), in which M antigen was heterozygous (Figure 1), but was detected by subsequent re-examination using a homozygous reagent. This indicates that blood group antigens having a dosage effect also affect detection sensitivity. In fact, it is reported that less sensitivity in automated detection of antibodies against antigens related to the Duffy and Kidd blood group systems compared to the tube technique seems to be related to the reagent’s zygotic status [3]. When screening clinically important antibodies, corresponding homozygous red blood cells need to be included in screening panels. One anti-Lea and two anti-M currently used for comparison of the detection sensitivity among three methods were not able to distinguish whether they were IgG-type or IgM-type due to weak agglutination reactivity. Though samples used in this study had been previously positive by PEG-IAT, repeated freezing and thawing might influence the reactivity.

The detection sensitivity of MP was higher than PEG-IAT in this study, but there were antibodies negative by MP which showed weak agglutination reaction by PEG-IAT. This phenomenon implies generally high sensitivity of the MP method, but it cannot detect all clinically important antibodies.

The mixed-field agglutination, or red cell chimera, results of this study clerly showed that tube technique was the most sensitive in discovering mixed chimeras composed of A1 and O cells, maybe because visual confirmation can easily distinguish double populations. However, automated MP and GC may not be adequate for clinical samples such as post-ABO mismatched hematopoietic stem cell transplantation, and post-ABO non-identical red cell transfusion for emergency.


CONCLUSION

Two automated systems investigated in this report concord well with tube technique in ABO forward and RhD typing. However, they appeared to be less sensitive than tube technique in detecting isohemagglutinin anti-A/anti-B in ABO reverse testing and in identifying ABO chimeras. Automation is efficient and can prevent some kinds of human errors, but in regard to the detection of low-titer, clinically significant antibodies, GC may fail to detect those that can endanger patients with post-transfusion hemolysis.


REFERENCES
  1. Morelati F, Revelli N, Maffei LM, et al. Evaluation of a new automated instrument for pretransfusion testing. Transfusion 1998 Oct;38(10):959–65.   [CrossRef]   [Pubmed]    Back to citation no. 1
  2. Sandler SG, Langeberg A, Avery N, Mintz PD. A fully automated blood typing system for hospital transfusion services. ABS2000 study group. Transfusion 2000 Feb;40(2):201–7.   [CrossRef]   [Pubmed]    Back to citation no. 2
  3. Weisbach V, Ziener A, Zimmermann R, Glaser A, Zingsem J, Eckstein R. Comparison of the performance of four microtube column agglutination systems in the detection of red cell alloantibodies. Transfusion 1999 Oct;39(10):1045–50.   [CrossRef]   [Pubmed]    Back to citation no. 3
  4. Hazenberg CA, Mulder MB, Beelen JM. Erythrocyte antibody screening in solid-phase: A comparison of two solid-phase microplate assays with the indirect antiglobulin test in polyethylene glycol for the detection of irregular erythrocyte antibodies. Vox Sang 1990;59(2):96–100.   [CrossRef]   [Pubmed]    Back to citation no. 4
  5. Winters JL, Richa EM, Bryant SC, Tauscher CD, Bendix BJ, Stubbs JR. Polyethylene glycol antiglobulin tube versus gel microcolumn: Influence on the incidence of delayed hemolytic transfusion reactions and delayed serologic transfusion reactions. Transfusion 2010 Jul;50(7):1444–52.   [CrossRef]   [Pubmed]    Back to citation no. 5
  6. Knowles SM, Milkins CE, Chapman JF, Scott M. The United Kingdom national external quality assessment scheme (blood transfusion laboratory practice): Trends in proficiency and practice between 1985 and 2000. Transfus Med 2002 Feb;12(1):11–23.   [CrossRef]   [Pubmed]    Back to citation no. 6
  7. Kay B, Poisson JL, Tuma CW, Shulman IA. Anti-Jka that are detected by solid-phase red blood cell adherence but missed by gel testing can cause hemolytic transfusion reactions. Transfusion 2016 Dec;56(12):2973–9.   [CrossRef]   [Pubmed]    Back to citation no. 7
  8. Okutsu M, Ohto H, Yasuda H, et al. Increased detection of clinically significant antibodies and decreased incidence of delayed haemolytic transfusion reaction with the indirect antiglobulin test potentiated by polyethylene glycol compared to albumin: A Japanese study. Blood Transfus 2011 Jul;9(3):311–9.   [CrossRef]   [Pubmed]    Back to citation no. 8
  9. Roback JD, Combs MR, Grossman BJ, Hillyer CD. Technical Manual. 16ed. Maryland: American Association of Blood Banks; 2008. p. 867–76.    Back to citation no. 9
  10. Ono T, Hikichi R, Kawabata K, Ohto H. Detection sensitivity of red cell alloantibodies using microtube column agglutination systems. Intern J Blood Transfus Immunohematol 2016;6:1–2.    Back to citation no. 10
  11. Langston MM, Procter JL, Cipolone KM, Stroncek DF. Evaluation of the gel system for ABO grouping and D typing. Transfusion 1999 Mar;39(3):300–5.   [Pubmed]    Back to citation no. 11
  12. Reid ME, Lomas-Francis C. The Blood Group Antigen Facts Book. New York: Academic Press; 1997. p. 19–26.    Back to citation no. 12
  13. Sazama K. Reports of 355 transfusion-associated deaths: 1976 through 1985. Transfusion 1990 Sep;30(7):583–90.   [Pubmed]    Back to citation no. 13
  14. Rolih S, Fisher F, Figueroa D, Lindsay G. Detection of tube agglutination 37 degrees C-only antibodies by solid-phase red cell adherence. Immunohematology 1996;12(1):27–9.   [Pubmed]    Back to citation no. 14
  15. Rolih S, Thomas R, Sinor L. Detection of Lewis, P1, and some MNS blood group system antibodies by a solid phase assay. Immunohematology 1995;11(3):78–80.   [Pubmed]    Back to citation no. 15
  16. Yamada C, Serrano-Rahman L, Vasovic LV, Mohandas K, Uehlinger J. Antibody identification using both automated solid-phase red cell adherence assay and a tube polyethylene glycol antiglobulin method. Transfusion 2008 Aug;48(8):1693–8.   [CrossRef]   [Pubmed]    Back to citation no. 16
  17. Villa MA, Moulds M, Coluccio EB, et al. An acute haemolytic transfusion reaction due to anti-Jka. Blood Transfus 2007 Apr;5(2):102–6.   [CrossRef]   [Pubmed]    Back to citation no. 17


[HTML Abstract]   [PDF Full Text]

Author Contributions
Takako Ono – Acquisition of data, Analysis and interpretation of data, Drafting the article, Revising it critically for important intellectual content, Final approval of the version to be published
Hitoshi Ohto – Substantial contribution to conception and design, Drafting the article, Revising it critically for important intellectual content, Final approval of the version to be published
Hiroyasu Yasuda – Substantial contribution to conception and design, Acquisition of data, Drafting the article, Revising it critically for important intellectual content, Final approval of the version to be published
Rie Hikichi – Acquisition of data, Drafting the article, Revising it critically for important intellectual content, Final approval of the version to be published
Kinuyo Kawabata – Substantial contribution to conception and design, Acquisition of data, Drafting the article, Revising it critically for important intellectual content, Final approval of the version to be published
Keiji Minakawa – Acquisition of data, Drafting the article, revising it critically for important intellectual content, Final approval of the version to be published
Satoshi Ono – Acquisition of data, Drafting the article, Revising it critically for important intellectual content, Final approval of the version to be published
Masami Kikuchi – Acquisition of data, Drafting the article, Revising it critically for important intellectual content, Final approval of the version to be published
Akiko Sugawara – Acquisition of data, Drafting the article, Revising it critically for important intellectual content, Final approval of the version to be published
Shunnichi Saito – Acquisition of data, Drafting the article, Revising it critically for important intellectual content, Final approval of the version to be published
Nozomi Takano – Acquisition of data, Drafting the article, Revising it critically for important intellectual content, Final approval of the version to be published
Kenneth E. Nollet – Drafting the article, Revising it critically for important intellectual content, Final approval of the version to be published
Guarantor of submission
The corresponding author is the guarantor of submission.
Source of support
None
Conflict of interest
Authors declare no conflict of interest.
Copyright
© 2017 Takako Ono et al. This article is distributed under the terms of Creative Commons Attribution License which permits unrestricted use, distribution and reproduction in any medium provided the original author(s) and original publisher are properly credited. Please see the copyright policy on the journal website for more information.

About the Authors

Takako Ono served as a medical technologist in Fukushima Medical University’s Department of Blood Transfusion and Transplantation Immunology. She graduated from Fukushima Prefectural General Hygiene Institute. Her main research interests lie in hematology, and in laboratory techniques for transfusion and transplantation. She is now pursuing advanced training in the United States.
Email: takako.ono.medtech@gmail.com


Hitoshi Ohto MD, PhD, is General Vice President of Fukushima Medical University, Japan. He has published more than 500 research papers in Japanese and international academic journals and authored 100 books. In addition to transfusion medicine, his research interests include disaster- and radiation-related issues.
Email: hit-ohto@fmu.ac.jp



Hiroyasu Yasuda graduated from Tohoku University and was the senior medical technologist in Fukushima Medical University’s Department of Blood Transfusion and Transplantation Immunology until shifting full-time into education at the Fukushima Prefectural General Hygiene Institute. His main research interests lie in immunohematology, and in laboratory techniques for transfusion and transplantation.
Email: hyasuda@fmu.ac.jp



Rie Hikichi graduated from Fukushima Prefectural General Hygiene Institute and is a medical technologist in the Department of Blood Transfusion and Transplantation Immunology at Fukushima Medical University in Fukushima, Japan. Her main research interests lie in hematology, and in laboratory techniques for transfusion and transplantation.



Kinuyo Kawabata graduated from Tohoku University and is a chief medical technologist in the Department of Blood Transfusion and Transplantation Immunology at Fukushima Medical University in Fukushima, Japan. Her main research interests lie in immunehematology, and in laboratory techniques for transfusion and transplantation.
Email: kawabata@fmu.ac.jp



Keiji Minakawa is a medical technologist in the Department of Blood Transfusion and Transplantation Immunology at Fukushima Medical University in Fukushima, Japan. He earned an undergraduate degree in health science from Tohoku University.
Email: m-keiji@fmu.ac.jp



Satoshi Ono is a medical technologist in the Department of Blood Transfusion and Transplantation Immunology, Fukushima Medical University Hospital, Fukushima, Japan.



Masami Kikuchi graduated from Fukushima Prefectural General Hygiene Institute and joined the Department of Blood Transfusion and Transplantation Immunology at Fukushima Medical University in Fukushima, Japan, as a medical technologist



Akiko Sugawara is a medical technologist in the Department of Blood Transfusion and Transplantation Immunology, and a research associate in the Center for Medical Education and Career Development at Fukushima Medical University, Fukushima, Japan. She received a PhD in Medicine from Fukushima Medical University. Her research interests include patient-provider communication, medical education, and transfusion medicine.



Shunnichi Saito graduated from Fukushima Prefectural General Hygiene Institute and served as a medical technologist in the Department of Blood Transfusion and Transplantation Immunology at Fukushima Medical University in Fukushima, Japan.



Nozomi Takano graduated from Fukushima Prefectural General Hygiene Institute and is a medical technologist in the Department of Blood Transfusion and Transplantation Immunology at Fukushima Medical University in Fukushima, Japan. Her main research interests lie in hematology, and in laboratory techniques for transfusion and transplantation.



Kenneth E. Nollet Professor, Department of Blood Transfusion and Transplantation Immunology, Fukushima Medical University Hospital, Fukushima, Japan.